How to solve the biochemical incubator without cooling

Shima Han biochemical incubator is suitable for environmental protection, sanitation and epidemic prevention, drug testing, agriculture, livestock, aquatic products and other research, colleges, production departments, water body analysis and 800 determination, bacteria, mold, microbial culture, preservation, plant cultivation, breeding experiments Dedicated thermostat.

In the process of using the island Korean biochemical incubator, we often encounter the situation that the incubator cannot be cooled. Here, the reasons and solutions are briefly explained:

(1) All operations of the incubator are effective, heating is normal, and cooling is not possible. Inspection method: First check whether the cooling indicator is on. If it is on, check whether the refrigeration relay has output. If there is no output, the contact will fall off or be damaged during frequent operation. Replace it. If there is output, check whether the compressor is running. If the refrigerant leaks during operation or the compressor cylinder wears out due to excessive use, there are many cases in the previous one. Solution: Find out the leaking fluorine point repair. The latter type only replaces the compressor (2) The incubator is effective for all operations, but the cooling indicator is not lit, the heating is normal, and the cooling cannot be performed. This situation is mostly caused by human error, just adjust the CL by entering the internal parameters.